Integrated Alarms for Patient Safety

CARINA Electronic Anaesthesia Ventilator
The OES Carina Anaesthesia Ventilator is an electronic time cycled ventilator with integrated patient alarms for improved patient safety
The Carina ventilator has adjustable inspiratory time, expiratory time and flow rate.
Adjustable low and high pressure alarms are provided, the low pressure alarm is the minimum patient circuit pressure that must be achieved with each breath while the high pressure alarm stops the inspiratory flow when the set high pressure is achieved.
The Carina can be used on paediatrics by changing the bellows unit.
The Carina can be sat on any flat surface or mounted on a dove tail bracket for use on the side of theĀ Astra Anaesthesia Unit machine
Specification | |
Inspiratory time | 0.25 – 3.0 seconds |
Expiratory time | 0.5 – 6.0 seconds |
Flow Rate | 0.1 – 1.0 litres per second |
Adjustable Pressure Relief | 20 – 80 cm H2O |
I:E Ratio | derived from inspiratory and expiratory time minimum 1:1 ratio |
Tidal Volume | 25ml – 1600ml Adult Bellows 25ml – 300ml Paediatric Bellows |
Driving Gas | Oxygen |
Supply Pressure | 45 – 100 PSI (310 – 700 KPA) |
Airway Pressure Guage | -20 – 100cm H2O Range |
Dimensions H x W x D (mm) | 340 x 300 x 260 mm |
Weight | 8 kg |
Adjustable Low Pressure Alarm | 0 – 30cm H2O |
Adjustable High Pressure Alarm | 10 – 70cm H2O |
Mute Button | with 15 second delay and LED |
Patient Audible and Visual Alarms | For disconnect / low airway pressure, high airway pressure, high continuous pressure and sub ambient pressure |
Additional Ventilator Alarms | For mains failure, low battery and low drive gas pressure |
External Power Supply | 90 – 264 volts A.C. |
Battery Backup | 12 hours minimum |
Ordering Information | |
9050 – 200 | Carina Anaesthesia Ventilator |
1988 – 503 | User Manual |
9070 – 001 | Adult Bellows Assembly |
9070 – 008 | Paediatric Bellows Assembly |
9070 – 003 | Adult Bellows Only |
9070 – 010 | Paediatric Bellows Only |
9060 – 010 | Bellows Drive Hose |
9060 – 012 | Patient Circuit – Drive Hose |
9060 -015 | Pressure Sensing Hose 2.5m and 22mm Male / Female Connector |
9050 – 017 | Anaesthesia Machine Dove Tail Bracket |